FOCAS Vacuum Pumping manual
Notes
CCDs should be operated under -100C. To avoid dew condensation within
the dewar containing CCDs and the electronics circuit, they should be kept
under high vacuum environment while the cooler is operating. Breaking the
vacuum under <-10C would seriously damage the CCD system.
An ion pump can be activated only under pre-vacuumed conditions, not under
ambient air pressure. The external pump will pump the dewar until the vacuum
level where the ion pump can be activated is reached.
Vacuum Layout
==; ||: vacuum tube
--; |: electric cable
[FOCAS INSIDE]
|--------------------|
|Dewar ==T connector==|== inner valve (yellow) == long tube == T connector == ion pump -- controller
| || | ||
| Vacuum gauge | outer valve (white) == external pump
|-----------|--------|
|
Vacuum gauge display
Procedures
Note: To read temperature, vacuum, etc., An AUX/NET breaker inside the FOCAS
main power distribution panel needs to be ON to activate the equipments.
Please turn it ON if it was set OFF. Do not touch other breakers unless you
are requested to do so.
A: Check the CCD temperature
If the CCD temperature is < 0 C, you cannot do pumping.
If the CCD temperature is close to the ambient temperature (~5C), proceed
to the followings.
B: Check the vacuum gauge
The vacuum level should be 1e-1 ~ 1 torr (the value may vary according
to the previous usage and conditions).
If the vacuum level is 1e-3 torr or less, you cannot do pumping.
C: Check the valves
There are two valves. Inner one (closer to the CCD chamber inside
the FOCAS at rear side at Cs) and the outer one (at front/IR side).
Left: Inner valve (a cylinder with a yellow screw), Right: Outer valve
(recently replaced to the white one. This picture shows the old
one.)
Close the inner one if it was open. Screw the yellow portion of the
valve completely to the end. Screw it gently.
D: Connect the external vacuum pump to the outlet at the outer valve
Use a portable VARIAN pump stored at the 3rd floor of the control
building.
Make sure that both o-ring and flanges are clean of oil/debris.
Clean them with alcohol and kim wipe if necessary, and grease
them up. Use as little vacuum grease as possible. Excess grease often works
to worsen the vacuum.
Fasten the vacuum clump securely
E: Open the outer valve
F: Start up an external pump
Make sure that the release valve (a small (~1 inch size) screw
at the side wall of the turbo pump) is closed.
Press START button on the pump controller.
Wait for a sign "normal operation 56K" shows up on the controller.
G: Open the inner valve
To avoid sudden change of the vacuum level in the CCD chamber,
open it very slowly while checking the vacuum gauge. Stop opening the valve
if the vacuum level rises in a short time scale (~several seconds) and
wait for a while (~30 seconds) until it stops rising.
Open it completely at the end.
H: Wait for 2-3 hours until it becomes less than 3E-4 torr
I: Start an ion pump
The ion pump controller is a box with a white panel showing some numbers.
- Set it to local mode (LCL)
- Switch a main power of the pump controller on.
- Switch a HV (high voltage) power on. Do not push HV power
on (lock). Switch it back off in 1 second.
- Check the vacuum gauge. Due to the out gas from the ion pump,
the vacuum quickly rises up. Wait for a minute until the vacuum gets lower
than the level at the start up.
- Switch the HV power back on again, and switch it back
off in 2 seconds. Wait for the pumping out of the out gas by the external
pump.
- Repeats the above sequence (HV on, off, wait) several ~ 10
times until the out gas is almost negligible while the HV is on. The
number of iterations required may vary according to the vacuum level at the
start up of this sequence.
- Check if the vacuum level is lower than the level at the beginning
of the ion pump start up at the end of the sequence.
J: Close the outer valve
Now the vacuum is maintained by only the ion pump. Make sure that
the vacuum level is almost constant even after closing the valve. If it raises
slowly, there must be something wrong (i.e., leak).
K: Disconnect the external pump
Stop the external pump by pushing the button on the panel of the
pump.
Open the release valve very (a small screw) very slowly. Take at least
one minute to open the screw completely. After the release, do not forget
to close it.
Cover the flanges with caps and clumps.
L: Set the ion pump controller to Remote
After checking everything is OK, the ion-pump controller should
be set to be Remote.
Call SA (he may be at the control room in the control building, or even
at Hilo/HP). This is a quick three step procedure.
- Set it remote on the controller front panel. The ion pump will be
temporarily OFF. All the LEDs (except for the main power indicator) will
go off.
- Ask SA to re-activate it remotely. He will turn it back ON using
the remote control software.
- Check the front panel to see the ion pump conditions. If all LEDs
go back on, it is OK.