FOCAS observations (and related telescope
operations) are made according to the following commands. These
commands are described in the operation procedure file (here for
more details).
Ask SA for POL/SPEC_POL commands
Command name |
Category |
Description |
Options |
Parameters |
Typical execution time |
Tips |
||
---|---|---|---|---|---|---|---|---|
Focusing
FOCAS |
FocusOBE | TOOL | Focusing FOCAS camera by way of Hartmann Test | ~10 min | Do CAL setup before executing this command. | Should be done once at evening. Normally made by the operator. | ||
Focusing Telescope |
FocusAGSuquence |
TOOL |
Telescope focusing with AG camera (including nearby bright star search) |
Z: Initial guess value of the secondary mirror position (Z) |
~3min |
Focus offset between AG and FOCAS will be applied automatically after finding the best Z position for AG. |
Focusing checks are suggested to be made at least twice a night (at evening and around midnight) during breaks between long exposures. | |
FocusAG |
TOOL |
Telescope focusing with AG camera |
Z: (see above) |
~2min | Same as FocusAGSequence, but without telescope pointing. |
If
you are NOT satisfied with the result made with
FocusAGSequence, use this command for iteration. |
||
FocusTelescope |
IMAG |
Focusing
telescope
directly with FOCAS |
Z: (see above) |
~10 min |
3X7 multiple point exposures will be made,
while changing telescope 2ndary mirror positions. |
Normally just FocusAG gives good result. |
||
Telescope Pointing |
Setupfield |
IMAG |
Pointing to
target. |
AG=YES: AG=NO: |
Insrot_PA: |
1min
+ telescope slewing (0.5deg/sec) |
|
No image is taken. If you need to check the image before going to long exposure, submit a short getobject command separately. |
SPEC |
Target
acquisition on a long slit |
AG=YES: AG=NO: If you need a specific filter for pointing images (2nd and 3rd steps, see note), select an appropriate filter. If the target is very faint, imaging without filter ($FILTER_NONE) gives good results. MODE=FINE: mode in case for another iteration for better accuracy. |
shift_sec:
(see above)
ExpTime: Exposure time for object
viewing
without slit (sec) (2nd and 3rd steps. see note) Insrot_PA: Slit PA (N=0, E=90) in deg. |
5-10min
+ telescope slewing (0.5deg/sec) |
Rough estimate for the exptime in case of point source:
Spec pointing toward EL<30 is prohibited, due to mechanical limitations of the slit exchanging system. |
If you want to adjust the slit PA precisely, or you do not know the accurate PA before the observation, ask SS for the detailed tips. In case of setupfield for the bright standard star, AG=NO works well. |
||
MOS |
Target
acquisition on
MOS mask |
MODE=FINE:(see above) |
Insrot_PA: (see above) |
20min
+ telescope slewing (0.5deg/sec) |
AG should be ON, and is default. ExpTime is 30 sec by default, and is good for most cases. MOS pointing toward EL<30 is prohibited,
due to mechanical limitations of the slit exchanging system. Require MOS pointing software running on ANA |
See
here for the detail
of
the MOS observation. |
||
Get Exposures |
getobject |
IMAG |
Image exposure |
ExpTime: exposure time |
Exposure time + CCD readout-time | |||
SPEC or MOS |
Spectroscopy exposure |
ExpTime: (see above) |
|
|||||
getstandard |
IMAG or SPEC |
Standard star exposure |
ExpTime: (see above) |
|
||||
Small Telescope Move |
movetelescope |
TOOL |
Telescope move in RA, DEC while not guiding |
delta_RA,_DEC: |
1-5sec |
|
||
moveguide0 |
TOOL |
Telescope move while guiding in RA, DEC |
delta_RA,_DEC:(see
above) |
5-10sec(depends on AG star mag) |
AG probe moves according to the telescope move. |
Good for larger distance move (>4") with less accuracy. | ||
moveguide3 |
TOOL |
Telescope move while guiding in RA, DEC |
delta_RA,_DEC:(see
above) |
5-10sec (depends on AG star mag) | Change of base point of the guiding. Shift values are calculated from current position. |
Good for shorter distance move (<4 ") with higher accuracy. | ||
moveguide |
TOOL |
Telescope move while guiding (short distance) at any PA. |
Slit_sec:
Movement along slit (arcsec)
Insrot_PA: Slit PA (N=0, E=90) |
5-10sec(depends on AG star mag) | Slit_sec > 4 arcsec: AG proves move according to the
telescope move, resulting in a slightly less pointing accuracy. Slit_sec < 4 arcsec: higher accuracy. |
Positive
Slit_sec value moves objects toward positive X
direction on chip, or toward chip 1. Insrot_PA should be set at same angle as that of target definition. |
||
Getting Bias | Getbias |
IMAG |
Taking Bias image |
~CCD readout-time | Use this command even for spectroscopic observation. |
Bias may not be taken frequently. Subtraction of the overscan value gives good results in many cases. | ||
Getting ARC image (wavelength calibration) |
SetupComparison |
TOOL |
Calibration unit
setup |
~1min | AG will go off while taking CAL lamp. You can re-start guiding with the same star after shutdonwncomparison is applied. |
Due to flexure, CCD position changes maximum 1 pix within 1 hour exposure (15deg shift of FOCAS orientation). If you require higher wavelength accuracy, take CAL every 1-1.5 hours. |
||
ShutdownComparison |
TOOL |
Calibration unit
shutdown |
|
~1min | Turn off the lamp, and retract the CAL probe. |
OH sky lines can be used as a good wavelength calibrator for red part of the spectrum, and can be used for correcting the wavelength drift among exposures. See here for the detail of the sky lines. |
||
GetComparison |
SPEC |
Arc spectroscopy for long slit |
~CCD readout-time | See here for exposure estimation |
You may skip the CAL observation if you use only OH lines. |
|||
MOS |
Arc
spectroscopy for MOS. |
Apply
three multi-exposure
while
sweeping the CAL probe for even illumination over the entire MOS mask. |
~CCD
readout-time + 2 min. |
To
obtain three multi-exposure while changing CAL probe positions, this
will take longer time than GetComparison for Spec. |
||||
GetComparison_MASKMOVE | MOS | Arc spectroscopy for MOS. | (see above) | ~CCD readout-time + 2 min. | This
is only for obtaining ARC spectra for MOS at the end of night, after
closing dome and telescope pointing to zenith. |
|||
Getting Dome Flat | GetDomeFlat |
IMAG |
Flat imaging |
VOLT: Lamp voltage Exptime: (sec) |
Exposure + CCD readout-time | See here for Exposure/Voltage estimation |
At evening or morning for IMAG and SPEC. Only at morning for MOS. |
|
Getting Dark Frames | GetDark |
IMAG |
Dark imaging |
Exposure + CCD readout-time | Dark is negligible in most cases, and you do not have take them. |