News
This Month
Introduction
Gallery
Science
Observing
|
Observation
Further information
Questions regarding this page should be directed to Naoyuki Tamura.
Planning observation
- Prepare a source list
Observers first need to make a list of coordinates for science
targets, guide stars, and coordinate calibration stars (see below
for details of the latter two). All the candidates can be included in
the list at this stage, since target selection for an optimal spine
allocation will be performed by a software (see the next section). To
ensure the accurate spine allocation and subsequently minimize flux
loss from the fibers, observers need to be careful about the following
things:
- Astrometry
Relative astrometry of your science targets compared with guide stars
and coordinate calibration stars is critical
because:
- The fibers are positioned to science targets by using guide
star positions as fiducials.
- Coordinate calibration stars are used to "align" the telescope
and instrument to the target field by applying lateral offset to
telescope pointing and rotational offset.
The error of spine allocation in a purely mechanical sense is ~10
μm (i.e. ~0.1 arcsec) and the error of relative astrometry should
not be much worse than this. In order to check the relative astrometry
of science targets, guide stars, and coordinate calibration stars,
observers need to cross-check the coordinates of stars on their
catalogs with those in a good astrometric catalog (e.g. UCAC3, 2MASS
point source catalog). UCAC3 would be particularly recommeneded
because stars are catalogued down to R ∼ 16 mag from the entire
sky, which is suitable for selection of guide stars and coordinate
calibration stars. - Please be aware that the cameras to
observe guide stars & coordinate calibration stars are CCD cameras for
visible wavelengths, while FMOS is a NIR instrument.
- Note:
-
The guideline of the brightness of guide star and coordinate
calibration star is give below in R band magnitude, since the sky
camera and guide camera are used with a red-pass filter
applied. Meanwhile, as explained in
this page, in
Section 3 (Properties of the catalog and important notes for the
user), Item d (Magnitudes), the "R-band" magnitude in the UCAC3
catalog is not that in the standard R band but in a 579-642 nm
bandpass which is between V and R. Observers need to take this into
account in selecting stars from the UCAC3 catalog based on the
"R-band" magnitude.
- Guide stars (GS)
Auto guiding (AG) is performed by taking images of the guide fiber
bundles (GFB) and calculating the centroids of GS. In total there are
14 GFB located at the edge of the field of view (FOV) as shown in
Fig. 1a (NOTE: Two GFB are currently unavailable due to their less
good positioning cababilities). The AG camera is a CCD camera with
a long-pass filter (OG590, transparent at > 0.6 μm) in front of
it. Observers therefore need to have GS bright in red optical bands
and located near the edge of FOV. A guideline for GS brightness
is R = 12 - 16 mag (this of course
depends on seeing condition and sky transparency).
Observers are highly recommneded to choose field center and PA to
maximize the number of available guide stars:
The more guide stars are available, the more stable AG operation is
expected. Especially, the operation should be rubust against
unexpected contamination of stars with large proper motions, double
stars, those with large errors in astrometry, magnitude, etc. In
addition, if all GS are bright, the exposure time for AG could be
shortened. Then telescope pointing correction could be applied at
a shorter interval and subsequently guding error could be smaller.
 |
|
|
|
 |
Fig. 1a - A sketch of the guide fiber
bundles. Note that two guide fiber bundles are unavailable due to
their less good positioning cababilities. |
|
|
|
Fig. 1b - A collection of the images of
14 guide fiber bundles during an actual auto guiding. 9 guide stars
are clearly visible around the center of each bundle. |
- Coodinate calibration stars (CCS)
These stars are used to align the telescope and instrument to the
target field and acquire GS on GFB, by applying corrections to the
telescope pointing and/or instrument rotator angle. This alignment (or
field acquisition) process is necessary because:
- The typical pointing error of the telescope (≈ 4-5
arcsec) is larger than the FOV of GFB (≈ 2.5 arcsec in
diameter), GS are not necessarily acquired by GFB after telescope
pointing.
- If a rotational offset remains, it cannot be fully corrected by
auto guiding, resulting in a significant amount of flux loss of
science targets.
To do this alignment, the sky camera (a CCD camera in the Echidna
Focal Plane Imager with an FOV of ~ 1 arcmin, with a red-pass filter
OG590) moves to the positions of CCS in the FMOS FOV and takes
images of CCS to measure the displacements of their positions from
where they should be.
A guideline for CCS brightness is R = 12 - 15
mag.
To work out not only lateral offset but also rotational offset
simultaneously, observers need to select one CCS near the field center
and a few stars surrounding it in the outer field.
NOTE(1): The CCS near the center of FOV is also used for focusing
operation: The same magnitude limit as other CCS (R = 12 - 15 mag)
is applied. If there are CCS only at distances larger than 10 arcmin
from the center of the FOV and one of those stars has to be used for
focusing operation, the focus value derived could be less accurate due
to the effect of curvature of focal plane.
NOTE(2): Observers may leave many CCS available in S2O files
(see below for details about S2O file): Although the actual field
acquisition is performed by using ∼ 5 CCS, if more CCS are
available in an S2O file then the instrument control system picks up
several from them considering the positions in the FoV.
- Run the spine allocation software
!! Important notes for S11B observers !!
- The Spine-to-object (S2O) allocation software is still updated
from time to time for bug-fix and upgrade. The latest version has
been ready to try since Jun 30, 2011. However, due to the recent
incident of the telescope, unfortunately, the engineering
observations in July and August were both canceled, and we have not
been able to confirm yet if S2O files created by the latest version
work in actual observation.
Accordingly, we request S11B observers to download an older version
(20101007 version) from the link below and prepare s2o files using
it. In fact, there are several known issues on this version, but
workarounds are available for most of them. S2O files created by
this version have been working in observations.
Note that there will be an engineering observation in Sep 2011, so
the latest version of S2O software may be available for observers
whose observations have been scheduled later in S11B.
- In S11B, HR as well as LR will be available on IRS1, while IRS2
will still be operated only in LR.
|
Observers will need to prepare and provide the information as below
using this S2O software:
- Field center
- Position angle
- Observing wavelength
Due to the effects of atmospheric dispersion and chromatic
aberration, the position of an object on the focal plane is
slightly different as a function of wavelength. Observers
therefore need to rightly input "observing wavelength" in the
S2O software. Below listed is the guideline of observing
wavelength for each observation mode:
Observing mode |
Observing wavelength |
Low resolution |
1.3-1.4 μm |
High resolution |
J-short |
1.0 μm |
J-long |
1.20-1.25 μm |
H-short |
1.50-1.55 μm |
H-long |
1.65-1.70 μm |
- Observation date
- Maximal hour angle acceptable for observation
- Source list (priorities can be given to the science targets)
- RA & DEC offsets for beam switching (unless the observing method is Point & Stare - see below)
- Observing method
- Normal beam switching --- In this mode, the
telescope is offset between "ON" and "OFF" positions, where
the fibers look at objects (sky) at the "ON" ("OFF")
position, respectively. Half of the observation time will be
spent for sky exposure.
- Cross beam switching --- In this mode, two
fibers are allocated to one object, and the telescope is
offset between two positions so that either of the two
fibers observes the object and the other observes sky. The
advantages of this method are: (1) 100% of the time can be
spent observing objects, and (2) at least in principle, sky
subtraction is not affected by time variation of sky
brightness. The disadvantage is that the maximal number of
spines allocated to objects is 200. Since the geometrical
constraint to spine allocation is strong, the actual number
of allocated spines could be even smaller in
reality.
- Point & stare --- There is no telescope offset
in this mode. Instead, some fibers need to be placed on
blank sky region and the average of the sky spectra is
applied to other fibers for sky subtraction. For example,
when most of the objects are relatively bright and the
accuracy of sky subtraction is not extremely critical, this
mode can be used.
The software then derives an optimal spine allocation and position
angle of the instrument considering a number of matters such as:
- Maximizing the number of high priority targets with spines allocated
- Maximizing the number of guide stars located on the guide fiber bundles
- Minimizing spine tilts
- Avoiding spines to cross each other

Fig. 2 - A snapshot of the spine allocation software. The arrows
show the spines allocated to science targets or guide stars.
- Save the output file(s) (*.s2o) and send it to the contact
person(s) for FMOS open-use observation.
The spine allocation software produces an output file (*.s2o).
According to the information in this S2O file, the Echidna Instrument
Control Software (ICS) will configure the positions of the science
fibers and guide fiber bundles for actual observations.
To finalize the spine allocation, please refer to the items in
this check list and check if
everything is considered properly.
Observers should send the final version of S2O files to support
astronomer(s) for FMOS open-use observations
(Kentaro Aoki ). If you observe more
than one spine allocations/target fields, please send all the
corresponding S2O files.
For clarity and maintenance purpose, please name your S2O files
following the format
"PROPOSALID_PILASTNAME_ANYCOMMENT_OBS.s2o". In the
"ANYCOMMENT" part, observers can choose a useful phrase to
classify the file (e.g. name of target field). "OBS" should be
replaced with one of "nbs", "cbs", or "pas" depending on the selected
observing method: Normal Beam Switching, Cross Beam Switching, or
Point & Stare, respectively (see above in this section for the details
of the observing methods). For example, if your proposal ID is
S10A-000 by PI: Tamura and you observe a target field named
"somewhere" by cross beam swiching, then please name the S2O file as
"s10a_000_tamura_somewhere_cbs.s2o".
NOTE: S2O file name needs to be CASE-INSENSITIVE.
Although it is OK to write an S2O file name using either all upper
cases or all lower cases, please DO NOT mix them. For
example, tamura.s2o and TAMURA.S2O are recognized as the same
file by the Echidna ICS.
- Prepare an OPE file
- What is OPE file?
Observation Procedure Execution (OPE) file contains a list of
abstract commands to operate the telescope and instrument in actual
observations. Like the other instuments on Subaru, observers using
FMOS will also need to prepare an OPE file according to their plans of
data acquisition during a night. Target information, observing method,
exposure time, and so on will also need to be specified in this
file.
- How is it used?
In an actual observation, the content of an OPE file is displayed on
the main telescope system control GUI. A support astronomer or
telescope operator then selects the line for a certain command in it
(with parameters in many cases) to execute when it is necessary.
Observations proceed usually by repeating this for different
commands.
Although an OPE file can be edited on the GUI during a night, to
minimize overhead, it is important for observers to provide all the
target information and prepare all the commands in OPE files that are
used in their observations. On the other hand, since a command is
manually picked up from an OPE file and is executed one by one, it is
not extremely important to organize the commands exactly along the
time sequence in an OPE file.
- Template OPE file
A template of OPE file for FMOS observation is available here:
Observers will need to download the text version of template and edit
it to make one for your observation. Please refer to the PDF version
as a brief manual. Then it needs to be sent to the support astronomer
for FMOS open-use observations (Kentaro Aoki ).
For clarity, please name your OPE file following the format
"PROPOSALID_PILASTNAME.ope". For example, if your proposal ID
is S10A-000 by PI: Tamura, then the name of an OPE file would be
"s10a_000_tamura.ope". Again the OPE file name needs to
be CASE-INSENSITIVE: Except for the extension (".ope", which
has to be lower case), either upper-case or lower-case characters can
be used for the file name but please do not mix them.
Observation procedure
- Pointing the telescope, rotating the instrument, configuring
spines, & checking the focus
- Target field acquisition by making corrections to telescope
pointing and instrument rotator angle
The typical overhead to complete 1. and 2. and get ready to start
exposures is ~20 minutes, which is usually dominated by the
spine configuration time. Spine configuration can start with slewing
the telescope and rotating the instrument. The overhead will be longer
when the telescope pointing is involved with a large amount of dome
rotation and/or when the next target field is still at EL<30 degree:
Since the spine configuration should be executed at a telescope
elevation NO LOWER THAN 30 degree, the telescope elevation is
temporarily raised to 40 degree to do a spine configuration.
- Start auto guiding, and take exposures
Notes for operation:
- Operations of IRS1 and IRS2 are independent. For example,
observers can set up IRS1 for LR and IRS2 for HR if wanted. Also, as
long as the telescope stays at a certain position and the fiber
configuration stays the same, exposures by IRS1 and IRS2 do not need
to be synchronized: Duration of individual exposure and number of
exposures can be chosen independently.
- As indicated in the "basic instrument
parameters" page, currently the readout noise of the detector in
the spectrographs is not very low. In a Correlated Double
Sampling (CDS) readout, this noise will have to be fully
included to every output frame. However, the noise level can be
significantly reduced by exploiting non-destructive readouts
(i.e. ramp sampling): Given an observer wants to take an 1800
sec exposure, the pixel count can be measured N=(1800/(minimum
exposure time)) times in the process of exposure and the final frame
can be obtained by a linear fit to the N data points. This is
expected to reduce the readout noise by a factor of sqrt(N).
Observers are therefore recommended to use this ramp sampling
unless they need to repeat very short exposure.
!! Important !! : Corrections to the spine
positions during long integration
Unfortunately, even if the instrument is rotated as necessary, the
positions of objects on the focal plane gradually change as time goes
by for several reasons as below:
- Since the rotator axis is slightly misaligned with the optical
axis of the telescope main mirror, the pattern of field distortion
on the focal plane rotates as the instrument rotates.
- Strength of differential atmospheric dispersion effect changes
as the telescope elevation changes.
- Plate scale changes, e.g., when the truss temperature changes,
and subsequently the distance between the telescope primary mirror
and wide-field corrector in the PIR changes.
Consequently, before flux loss from fibers starts being significant
due to a large displacement between fiber and object positions,
the spine positions need to be re-configured at a regular interval
to keep observing the same objects with the same spine
configuration. In the recent engineering observations, it has
been confirmed that re-configuring the spine positions every 30
minutes enables the object flux to be reasonably constant. Each
re-configuration process takes about 10 minutes. The typical
observation efficiency in long integration is ~60% (i.e. the overhead
is ~40%), while the efficiency gets lower for shorter integration
because the contribution of initial configuration time (20 minutes)
becomes more significant.
As of Feb 2011, we are still in the process of characterization and
optimization and this frequency of re-configurations may be reduced in
future as the instrument characteristics are better understood.
However, the observers are strongly recommeneded to follow this
sequence (i.e. executing the re-configuration process every 30
minutes) for long integration.
Data acquisition for flat fielding and calibration
- Flat fielding
Dome-flat frames are taken in the evening and/or morning by
closing the dome and using the dome-flat lamp. In doing this, the
Echidna spines are configured to the same as for a scientific
exposure. If observers have more than one target field/spine
configuration to observe, then they would need to take the same
number of sets of dome-flat frames with the spine configuration
changed as appropriately.
- Wavelength calibration
A Th/Ar lamp is available in the buffle structure of the
tertiary mirror pointing to the focal plane at the prime focus.
Using this lamp, CAL frames are taken in the evening and/or morning
(spine configuration at the time of this data acquisition should not
matter for the accuracy of the calibration). A line list will be
provided on this web site in the near future.
- Telluric absorption correction and flux
calibration
What has been usually done so far is to assign a few fibers to
observe faint stars simultaneously with science targets. A guideline
of the brightness of stars for this method is 15-18 mag (AB) in
JH. The brighter limit of this range is set to minimize the effect
of ghost features after the typical exposure time of an individual
frame (i.e. 15 min), which tends to appear at a
three-orders-of-magnitude (i.e. 7.5 mag) fainter level than the
original brightness. The fainter limit is set to keep S/N of these
stellar spectra high enough for calibration. The spectral types
preferred are F, G, and K early dwarfs (A stars can be handled by
the reduction package but are not recommeneded). Broad-band colors
are expected to be useful to select them in advance, while it is
also possible to estimate spectral type from observed spectrum,
given the instrument throughput.
In theory, such correction/calibration is possible if one star is
observed per spectrograph, but it is strongly recommended to observe
a few to several stars possibley of different types so that the
result of the correction/calibration can be cross-checked.
An alternative method is to observe a standard star in a different
field before and/or after observing a science target field.
Observers would need to prepare an S2O file for a standard star
observation separately from those for science target field
observations, where field center, CCS, & GS are necessary in the
same way as science fields. This method would be recommended
e.g. when a target field is at a low Galactic latitude and standard
stars available there as well as science targets are highly affected
by Galactic extinction.
Operation of high resolution mode
In the High Resolution mode (HR), the FMOS spectral coverage (0.9-1.8
μm) is divided into four pre-defined bands ("J-short", "J-long",
"H-short" & "H-long") with a band width of ~0.25 μm, and one of
them is observed at one exposure with a spectral resolution of ~5A.
Please check this page for the basic
parameters and this page for the
sensitivity information.
In S12A, HR as well as LR will be available on both IRS1 and
IRS2. There will be a few restrictions to the HR
operation as follows, and applicants/observers should keep
them in mind about the use of HR:
- The FMOS spectral coverage (0.9-1.8 um) is divided into four
bands in HR: J-short (0.90-1.16 μm), J-long (1.09-1.35
μm), H-short (1.40-1.66 μm) & H-long (1.54-1.80
μm), and one of them is covered by a single
exposure. HR observation is carried out by
using these pre-defined bands only. Note: More
flexible operation may be possible in future semesters, but as
of Aug 2011 more engineering works will be necessary to do so.
Hence we decided to apply the restriction in S12A (i.e. the
first semester of HR open-use) to minimize any technical risks.
- To minimize the risk of increasing additional overhead,
observers need to keep the same setting of IRS1 & IRS2 from
beginning to end of a night - i.e. No change
is allowed in observation mode (e.g. LR → HR, HR → HR
but a difference band) during a night.
- It would be acceptable to use IRS1 and IRS2 in different
observation modes (e.g. LR in IRS1 and J-long in IRS2),
while it is strongly recommended to use IRS1 and IRS2 in the
same mode to accurately position Echidna spines/fibers: Due
to the effects of atmospheric dispersion and chromatic
aberration, the position of an object on the focal plane is
slightly different as a function of wavelength, and the
spine-to-object (s2o) allocation software cannot allocate spines
according to different requests (e.g. observation wavelength) to
spines for IRS1 and those for IRS2. This means, if one tried
observation in J-long with IRS1 and in H-long with IRS2 and set
observation wavelength for spine allocation to 1.3 μm, more
light would tend to be lost at the fiber entrance for the H-long
observation.
|